products services downloads about ADV Contact Us home

 

service ranges...

The AMD-3875 12 Lead Interpretive ECG for PC combines unparalleled ease-of-use, flexibility and portability.  

The AMD-3875 converts any Windows 95/NT platform into a real time 12-lead ECG machine with Interpretation when connected to the serial port of a PC.

 

 

Dimensions:  1.3"(3.3 cm) H x 5"(12.7 cm) W x 7.10"(18.1 cm) L
Weight: 0.84 lbs. (.038kg)
Power / Electrical:  4 AA alkaline batteries
Intended Use:  Resting Electrocardiogram
Anatomical Sites:  Non-invasive device, 12-lead electrocardiogram
Safety Characteristics: 
  • Patient electrically isolated from mains.
  • Patient leakage current <10uA
  • Ground leakage current <50uA
ECG Acquisition:
  • Simultaneous 12-lead
  • Patient Connection: 10-lead patient cable with RFI filter, defibrillator protection and patient isolation.
  • Input impedence >100 Mohm
  • Frequency response 0.05-110 Hz
  • Sensitivity : 5, 10, 20mm/mV
  • ADC resolution 2.44uV/bit
Analysis & Measurement:
  • Telemed Electrocardiogram

 

 

The AMD-8200 Vital Signs Monitor provides a digital display of temperature, blood pressure, pulse and blood oximetry. 

The system can store up to 99 sets of data that can be seen on the monitor or downloaded. The monitor can be integrated with a PC via an RS-232 port. In addition, a visual and audible alarm can be set when blood pressure or pulse exceed set limits. 

It is also provided with SDK for easy integration into a patient record keeping system. The lightweight design and rechargeable battery allow for easy portability.

Dimensions:  6.5"(16.5 cm) H x 8.6"(21.8 cm) W x 5.0"(12.7 cm) L
Weight: 6 lbs. (2.8 kg)
Data Output:  RS-232
Blood Pressure Accuracy:  Meets or exceeds SP10-1992 AAMI standards
Cuff Pressure Range:  0-300 mmHg
Pressure Preset Levels:  120,140,160,200,240 or 280 mmHg
Systolic Determination:  Max: 250 mmHg; Min: 60 mmHg
Diastolic Determination:  Max: 160 mmHg; Min 30 mmHg
Pulse Rate Determination:  Max: 200 bpm; Min: 40 bpm
Temperature Accuracy:  Meets or exceeds ASTM E112-86 standards
Temperature Determination:  Normal and Monitor Mode, Max: 108.0 deg. F (42.2 deg. C); Min: 84.0 deg. F (28.9 deg. C)
Power Requirements:  International version: 220-240 VAC, 50-60 Hz
Internal Battery:  Lead Acid, with external recharger
Safety Standards:  IEC 60601-2-30, UL 2601-1, CAN/CSA C22.2 No.601.1
Conformity:  CE

 

For full details of the entire AMD range or to book 
yourself a demonstration please contact us.